Fluke 5720A Power Supply User Manual


 
Calibration and Verification
Determining Test Limits for Other Calibration Intervals
3
3-61
Table 3-24. 5700A AC Voltage Output Test Record (cont)
Output
Level
Frequency
Error
Display
Reading
Spec
(± ppm )
5790A
1-Year
Spec
(± ppm)
TUR GB
Factor
Test Limit (4)
(± ppm)
200V 1 kHz 85 31 2.74 0.971 83
200V 20 kHz 85 31 2.74 0.971 83
200V 50 kHz 240 69 3.48 0.991 238
200V 100 kHz 600 98 6.12 1.00
200V(2) 40 Hz 85 31 2.74 0.972 83
200V 20 Hz 185 68 2.72 0.970 179
200V 10 Hz 600 200 3.00 0.980 588
300V(3) 20 kHz 145 41 3.54 0.993 144
600V(3) 50 kHz 378 130 2.91 0.977 369
600V(3) 100 kHz 1375 500 2.21 0.945 1299
1 kV 1 kHz 84 38 2.21 0.945 79
1 kV 50 Hz 84 38 2.21 0.945 79
1 kV 300 Hz 84 38 2.21 0.945 79
1 kV(3) 20 kHz 131 38 3.45 0.991 130
1 kV(3) 30 kHz 371 130 2.85 0.975 362
1 kV(3) 40 Hz 84 38 2.21 0.945 79
1: This is a test of the bottom of the 20V range .
2: Observe the Calibrator output for 10 minutes and verify that it remains stable within
±
7.5 ppm.
3: Perform only for units that are used with a 5725A Amplifier.
4: When the TUR (Test Uncertainty Ratio) is less than 4:1, the spec is reduced to give the same
Consumer Risk as a 4:1 TUR as described later in this chapter.