Calibration and Verification
Performing a Calibration Check
3
3-21
PREV
MENU
IF3-3.EPS
3. Enter the ambient temperature; then press E. As the calibration check
proceeds, the display indicates the current process of the calibration check. When the
check is complete, the largest shift that is detected appears on the display.
PREV
MENU
IF3-17.EPS
4. You can now print a list of the proposed shifts, or quit without creating a list.
To print the report, connect a printer and set up the serial interface as described in
Chapter 6 of the 5700A/5720A Operators Manual. Press the softkey under “Print
Output Changes.” To return to normal operation without printing, press the softkey
under “DONE with check.”
Press P to exit the calibration menus.