Troubleshooting
Interpreting Diagnostic Fault Codes
5
5-39
3916 A5: 20DB Output Attenuation Fault
The 6.5V reference (BRF6, BSRF6) is connected to the Unity Gain Amplifier circuit on
the Wideband Output assembly which drives the output through the Attenuators. The
Wideband Output assembly is set for 20 dB attenuation so the 6.5V input results in a
0.325V output into 50. This 0.325V is connected to the RCL line which is routed to the
+input of the adc circuit on the DAC assembly. The DAC assembly output is connected
to the -input of the adc circuit and adjusted until a null is achieved. This determines the
exact value of the Wideband output. This fault occurs if the determined Wideband output
is over 6% ±1.0 mV. A failure indicates the Wideband Output (A5) assembly is probably
at fault.
3917 A5: 30DB Output Attenuation Fault
The 6.5V reference (BRF6, BSRF6) is connected to the Unity Gain Amplifier circuit on
the Wideband Output assembly which drives the output through the Attenuators. The
Wideband Output assembly is set for 30 dB attenuation so the 6.5V input results in a
0.103V output into 50. This 0.103V is connected to the RCL line which is routed to the
+input of the adc circuit on the DAC assembly. The DAC assembly output is connected
to the -input of the adc circuit and adjusted until a null is achieved. This determines the
exact value of the Wideband output. This fault occurs if the determined Wideband output
is over 6% ±1.0 mV. A failure indicates the Wideband Output (A5) assembly is probably
at fault.
3918 A5: 40DB Output Attenuation Fault
The 6.5V reference (BRF6, BSRF6) is connected to the Unity Gain Amplifier circuit on
the Wideband Output assembly which drives the output through the Attenuators. The
Wideband Output assembly is set for 40 dB attenuation so the 6.5V input results in a
0.032V output into 50. This 0.032V is connected to the RCL line which is routed to the
+input of the adc circuit on the DAC assembly. The DAC assembly output is connected
to the -input of the adc circuit and adjusted until a null is achieved. This determines the
exact value of the Wideband output. This fault occurs if the determined Wideband output
is over 6% ±1.0 mV. A failure indicates the Wideband Output (A5) assembly is probably
at fault.
3919 Spare
This fault code is not used.
3920 Spare
This fault code is not used.
3921 Spare
This fault code is not used.
3922 A5/A6: Wideband Initialization Fault
This fault occurs if the diagnostics software was unable to properly set the Wideband
assembly hardware for diagnostics testing. A failure indicates the Wideband Output (A5)
assembly is probably at fault.