Calibration and Verification
Full Verification
3
3-33
Table 3-8. 5790A Adjustment Counts
5700A/5720A Series II Output
Level
5790A Display 5790A Adjustment Counts*
2 mV 2.0000 ±1
20 mV 20.0000 ±1
200 mV 200.0000 ±4
2V 2.000000 ±4
2.3V 2.30000 ±1
20V 20.00000 ±4
200V 200.0000 ±4
300V 300.000 ±1
600V 600.000 ±2
1000V 1000.000 ±2
*Adjustment counts of 3 times the listed value is allowed at 1 MHz
3-23. AC Current Test
,
22 mA to 11A Ranges
This test requires the use of equipment listed in Table 3-9, when using the 5790A Input
1. Use Tables 3-33 (5720A) or 3-34 (5700A) for a test record.
Table 3-9. Equipment Required for 22 mA to 11A Alternating Current Test Using the 5790A Input 1
Equipment Model or Description
AC Measurement standard Fluke 5790A
Current Shunts Fluke A40 Series: 20 mA, 200 mA, 2A, and A40A Series: 10A (if verifying a
5725A) , with AC-DC difference corrections.
Current Shunt Adapter Fluke 792A-7004 A40 current shunt adapter
Cable Pomona 1368-A-18. Double banana to single banana plugs.
Cable (For 10A Setup) Pomona 5268-C-12. BNC(M) to single banana plugs.
Cable (For 10A Setup) Fluke A45-4003 (PN 212853) UHF(M) to UHF(M) with RG8A/U cable
Adapter (For 10A Setup) Pomona Model 1707. UHF(F) to banana adapter.
Adapter (For 10A Setup) Kings KC-99-34 UHF(M) to BNC(F)
Adapter (For 10A Setup) Pomona Model 1707. UHF(F) to banana adapter.