This will create a displayed list for sequence 1, as shown:
Start of Sequence
RECALL
FlEG
1
IF LIMIT TEST PASS THEN DO
SEQUENCE2
IF LIMIT TEST FAIL THEN DO
SEQUENCE3
2.
lb
create a sequence that stores the measurement data for a device that has passed the limit
test, press:
This will create a displayed list for sequence 2, as shown:
Start of Sequence
INTERNAL DISK
DATA ARRAY
ON
FILENAME
FILE0
SAVEFILE
3.
‘lb
create a sequence that prompts you to tune a device that has failed the limit test, and
calls sequence 1 to retest the device, press:
This will create a displayed list for sequence 3, as shown:
Start of Sequence
TITLE
TUNE DEVICE
SEQUENCE
PAUSE
DO SEQUENCE
SEQUENCE1
2-80
Making Measurements