HP (Hewlett-Packard) HP 8753E Network Cables User Manual


 
PLOT
ABORTED
Error Number When you press the
LLocal)
key, the analyzer aborts the plot in progress.
27
PLOTTER:noton,notconnect,
wrongaddrs
Error Number The plotter does not respond to control. Verify power to the plotter, and check
26
the HP-IB connection between the analyzer and the plotter. Ensure that the
plotter address recognized by the analyzer matches the HP-IB address set on
the plotter itself.
PLOTTERNOTREADY-PINCHWHEELSUP
Error Number The plotter pinch wheels clamp the paper in place. If you raise the pinch
28
wheels, the plotter indicates a “not ready” status on the bus
POSSIBLEFALSELOCK
Error Number Phase lock has been achieved, but the source may be phase locked to the
6
wrong harmonic of the synthesizer. Perform the source pretune correction
routine documented in the “Adjustments and Correction Constants” chapter in
the
HP
875333
Network Analyzer
semrice
Guide.
POWERMETERINVALID
Error Number The power meter indicates an out-of-range condition. Check the test setup.
116
POWERMETERNOT
SETTLED
Error Number Sequential power meter readings are not consistent. Verify that the equipment
118
is set up correctly. If so, preset the instrument and restart the operation.
POWERSUPPLYHOT!
Error Number The temperature sensors on the
A8
post-regulator assembly have detected an
21
over-temperature condition. The power supplies regulated on the
post-regulator have been shut down.
Error Messages
10-18