HP (Hewlett-Packard) HP 8753E Network Cables User Manual


 
SEGMENT#nPOUEROUTSIDERANGELIMIT
Information
The selected power range does not support the power level of one or more
Message
segments in the swept list table. This message appears when swept list mode is
not on and reports the
llrst segment that is out of range. Change the segment
power or change the power range.
SEGMENT#nSTARTFREQ
OVERLAPSPREVIOUSSEGMENT
Information
A segment entered in the swept list table caused one or more frequency
Message
segments to overlap. This message appears when swept list mode is not on and
reports the
first
segment that is overlapping another. Change the frequency
ranges of the overlapping segments.
SELECTEDSEQUENCEISENPTY
Error Number The sequence you attempted to run does not contain instnunent commands
124
SELFTESTkFAILED
Service Error Internal test #n has failed. Several internal test routines are executed at
Number 112
instrument preset. The analyzer reports the iirst failure detected. Refer to the
HP 8753E Network
Amlgm
Semrice
G&de for troubleshooting information on
internal tests and the self-diagnose feature.
SEQUENCE
ABORTED
Error Number The sequence running was stopped prematurely when you pressed the
llocalJ
157
key.
SEQUENCEMAYHAVECHANGED,CAN'TCONTINUE
Error Number When you pause a sequence, you cannot continue it if you have modified it.
153
You must start the sequence again.
Error Messages
1 O-23