Example
4,861O
S-Term
Frequency List Cal Set
F’ile
Example 4 shows how
CITIfIle
may be used to store instrument setup information. In the case
of an 8510
Cal
Set, a limited instrument state is needed in order to return the instrument to the
same state that it was in when the calibration was done.
Three arrays of error correction data are dellned by using three DATA statements. Some
instruments require these arrays to be in the proper order, from El to
E3.
In general,
CITIfile
implementations should strive to handle data arrays that are arranged in any order.
Example
CITIFILE
A.O1.OO
#NA
VERSION
HP8510B.05.00
NAME CAL-SET
#NA
REGISTER 1
VAR FREQ
MAG 4
DATA
EC11
RI
DATA
E[2]
RI
DATA
EC31
RI
#NA SWEEP-TIME
9.9999873-2
#NA
POWER1
l.OEl
#NA
POWER2
l.OEl
#NA
PARAMS
2
#NA
CAL-TYPE 3
#NA
POWER-SLOPE
O.OEO
tNA
SLOPE-MODE 0
#NA
TRIM-SWEEP 0
#NA
SWFaEP,MODE
4
#NA LOWPASS,FLAG -1
#NA
FREQJNFO
1
XNA SPAN 1000000000 3000000000 4
#NA DUPLICATES 0
#NA ARB,SEG 1000000000 1000000000
1
#NA
ARB,SEG 2000000000 3000000000 3
VAR,LIST-BEGIN
1000000000
2000000000
2500000000
3000000000
VAR,LIST-END
Continued next page