HP (Hewlett-Packard) HP 8753E Network Cables User Manual


 
PROBEPOWERSHUT
DOWN!
Error Number One or both of the probe power supplies have been shut down due to an
23
over-current, over-voltage, or under-voltage condition.
PROCESSINGDISPLAYLIST
Information The display information is being processed for a screen print to a copy device
Message
and stored in the copy spool buffer. During this time, the analyzer’s resources
are dedicated to this task (which takes less than a few seconds).
PWRMTRNOTON/CONNECTED
ORWRONGADDRS
Error Number The power meter cannot be accessed by the analyzer. Verify that the power
117
meter address and model number set in the analyzer match the address and
model number of the actual power meter.
RANGECAUSEDPOWERLVLCHANGEINLIST
Error Number The selected power range changed the power level of one or more segments in
213
the swept list table. Change the segment power or change the power range.
REQUESTEDDATANOTCURRENTLYAVAILABLE
Error Number The
analyzer
does not currently contain the data you have requested. For
30
example, this condition occurs when you request error term arrays and no
calibration is active.
SAVEFAILED.
INSUFFICIENTMEMORY
Error Number You cannot store an instrument state in an internal register due to insufficient
151
memory. Increase the available memory by clearing one or more
save/recall
registers and pressing
B,
or by storing files to a disk.
1 O-22