HP (Hewlett-Packard) HP 8753E Network Cables User Manual


 
NETWORK ANALYZER
PORl
ES
OPEN SHORT
LOAG
OPEN SHORT LOAD
FOR
S,,
FOP
Sz2
pgE14e
Figure 5-6. Standard Connections for a One Port Reflection Error-Correction
8. To measure the standard, when the displayed trace has settled, press:
./
g#&
:....
Note
If the calibration kit that you selected has a choice between male or female
calibration standards, remember to select the sex that applies to the test port
and not the standard.
The analyzer displays WAIT
-
MEASUFUNG
CAL STANDARD during the standard measurement.
The analyzer underlines the
&!J@#
softkey
after it measures the calibration standard.
9. Disconnect the open, and connect a short circuit to the test port.
10.
lb
measure the standard when the displayed trace has settled, press:
z#$#g!g
,.,)
;::
_
:.:.:y:<.
.,,,
&:.:::
.,/
_,
The analyzer measures the short circuit and underlines the
@J@RZ’
softkey.
*.
.._..
_..._
11. Disconnect the short, and connect an impedance-matched load to the test port.
12.
When
the
displayed
trace
settles,
press
~~~.
_
,. ,. .,
.
.
.
.
.
The analyzer measures the load and underlines the
&Q&j&
softkey.
.
13.
‘lb
compute the error coefficients, press:
The analyzer displays the corrected data trace. The analyzer also shows the notation Cor to
the left of the screen, indicating that the correction is switched on for this channel.
Note
The open, short, and load could be measured in any order, and need not follow
the order in this example.
Optimizing
Measurement
Results 5-19