Response Error-Correction for Transmission Measurements
1. Press
LPresetJ.
2. Select the type of measurement you want to make.
q If you want to make a transmission measurement in the forward direction
(&),
press:
q If you want to make a transmission measurement in the reverse direction
(&a),
press:
3. Set any other measurement parameters that you want for the device measurement: power,
number of points, IF bandwidth.
4.
To
select a response correction, press:
5. Make a “thru” connection between the points where you will connect your device under
test.
Note
Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device where you will connect your device under
test.
NETWORK ANALYZER
TEST PORT
POSSIBLE
ADAPTERS
pg611e
Figure 5-2.
Standard Connections for Response Error-Correction for Transmission Measurements
6.
‘Ib
measure the standard, press:
_
_
_
_
;
,...
/
_
;2J$!B&
.
The analyzer displays
WAIT
-
MEASURING CAL STANDARD
during the standard measurement.
The
analyzer
underlines the
~~
softkey after it measures the calibration standard, and
-
.._...._........
computes the error coefficients.
Optimizinfl
Measurement Results
5-l
1