HP (Hewlett-Packard) HP 8753E Network Cables User Manual


 
CAN'TSTORE/LOADSEQUENCE,
INSUFFICIENTMEMORY
Error Number Your sequence transfer to or from a disk could not be completed due to
127
insufficient memory.
CAUTION:
AUXCHANNELSMEASURES-PARAMETERS
ONLY
Error Number This message appears if you try to assign a non-S-parameter measurement to an
216
auxiliary channel.
CAUTION:
CORRECTIONOFF:
AUXCHANNEL(S)
DISABLED
Error Number This message is displayed when correction is forced off due to a stimulus
215
change that is not compatible with the current calibration while an auxiliary
channel is enabled. The auxiliary channels are restored when correction is
turned
on
by
pres,&,g
Lcal]
~~~~~~~~~~~.
CHl
(CH2)
TARGETVALUENOTFOUND
Error Number Your target value for the marker search function does not exist on the current
159
data trace.
CONTINUOUS
SWITCHINGNOTALLOWED
Error Number Your current measurement requires different power ranges on channel 1 and
channel 2.
To
protect the attenuator from undue mechanical wear, test set hold
10
will be activated.
The
“tsH”
(test set hold) indicator in the left margin of the display indicates
that the inactive channel has been put in the sweep hold mode.
COPY:
devicenotresponding;
copy
aborted
Error Number The printer or plotter is not accepting data. Verify the cable connections, HP-IB
170
addresses, and otherwise ensure that the copy device is ready.
COPYOUTPUTCOMPLETED
Information
The analyzer has completed outputting data to the printer or plotter. The
Message
analyzer can now accept another copy command.
1 O-6
Error
Messames