HP (Hewlett-Packard) HP 8753E Network Cables User Manual


 
Verification
Kit
Accurate operation of the analyzer system can be verified by measuring known devices other
than the standards used in calibration, and comparing the results with recorded data.
HP
86029B
7-mm
Verifkation
Kit
This kit contains traceable precision
7-mm
devices used to
confIrm
the system’s error-corrected
measurement uncertainty performance.
Also
included is verification data on a 3.5 inch disk,
together with a hard-copy listing. A system verification procedure is provided with this kit and
alsointhe
HP8753EServiceGuide.
Test
Port
Return Cables
The
foIIowing
RF cables are used to connect a two-port device between the test ports. These
cables provide shielding for high dynamic range measurements.
HP
11867D
7-mm
l&t
Port Return Cable Set
This set consists of a pair of test port return cables that can be used in measurements of
7-mm
devices. They can
also
be used with connectors other than 7 mm by using the appropriate
precision adapters.
HP
11867B
75 Ohm Type-N
Test
Port Return Cable Set
This set consists of test port return cables for use with the HP 8753E Option 075.
Adapter Kits
HP
11862B
50 to 75 Ohm Minimum Loss Pad.
This device converts impedance from 50 ohms to 75 ohms or from 75 ohms to 50 ohms It is
used to provide a low SWR impedance match between a 75 ohm device under test and the
HP 8753E network analyzer (without Option 075).
These adapter kits contain the connection hardware required for making measurements on
devices of the indicated connector type.
n
HP
11853A
50 Ohm Type-N Adapter Kit
n
HP
11854A
50 Ohm BNC Adapter Kit
n
HP
11855A
75 Ohm Type-N Adapter Kit
w
HP
11856A
75 Ohm BNC Adapter Kit
1
l-2
Compatible Peripherals