HP (Hewlett-Packard) HP 8753E Network Cables User Manual


 
NETWORK ANALYZER
FOR
S,,
RESPONSE FOR
S22
RESPONSE
SHORT OPEN
FOR
S,,
,SOLATlON
b
LOAD LOAD
pg612e
Figure 5-4.
Standard Connections for a Response and Isolation Error-Correction for Reflection
Measurements
8. To measure the standard, press:
If the calibration kit you selected has a choice between
male
and female calibration
standards, remember to select the sex that applies to the test port and not the standard.
The analyzer displays WAIT
-
MFASURING
CAL STANDARD during the standard measurement.
The analyzer underlines the
softkey
that you selected after it
finishes
the measurement,
and computes the error coefficients
9. Connect the load calibration standard to the test port.
10.
lb
measure the standard for the isolation portion of the correction, press:
11. To compute the response and directivity error coefficients, press:
The analyzer displays the corrected
SH
(or
S&
data. The analyzer also shows the notation
Cor to the left of the screen, indicating that the correction is switched on for this channel.
Note
You can save or store the error-correction to use for later measurements.
Refer to the “Printing, Plotting, and Saving Measurement Results” chapter for
procedures.
12. This completes the response and isolation error-correction for reflection measurements. You
can connect and measure your device under test.
Optimizing Measurement Results 5-15