CALCulate Subsystem
TVS600 & TVS600A Command Reference
2–69
CALCulate:WMParameter:HIGH
CALCulate:WMParameter:HIGH?
Sets or queries the high (most positive) level used for time and amplitude
measurements. This setting is used only when CALC:WMP:HMEThod is set to
ABSolute. For more information, refer to the discussion of CALC:WMP:HME-
Thod. The units are those currently used for vertical magnitude.
CALCulate<n>:WMParameter:HIGH <high>
CALCulate<n>:WMParameter:HIGH?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
<high> (Absolute HIGH level) Query response
<NRf> <NR3>
0.0E+0
Execution Error –222, “Data out of range”
Attempted to set HIGH to an illegal value.
None
Command: CALC1:WMP:HIGH 10.0
Query: CALC1:WMP:HIGH?
Response: 10.0E+0
CALCulate:WMParameter:HMEThod
CALCulate:WMParameter:LOW
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands