SYSTem:COMMunicate:SERial:PARity
SYSTem:COMMunicate:SERial:SBITs
SYSTem:COMMunicate:SERial:CONTrol:DCD
SYSTem:COMMunicate:SERial:CONTrol:DCD?
Sets or queries whether the instrument is sensitive to the front panel RS-232
DCD line. The setting ON specifies that the port ignores data unless the DCD
line is high. This mode is useful when you have a modem connected. The setting
OFF means the DCD line is ignored. Refer to the TVS600 & TVS600A Series
Waveform Analyzers User Manual for the factory setting.
When you set :SERial:CONTrol:DCD, the new value is stored in nonvolatile
RAM and is recalled at power up. However, the control mode is not saved with
stored settings (*LRN) and it is not reset to a default value by *RST.
SYSTem:COMMunicate:SERial:CONTrol:DCD <boolean>
SYSTem:COMMunicate:SERial:CONTrol:DCD?
<boolean> Query response
<NRf>
1 or ON
0 or OFF
<NR1>
1
0
Not applicable
None
None
Command: SYST:COMM:SER:CONT:DCD ON
Query: SYST:COMM:SER:CONT:DCD?