Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
INPut Subsystem
TVS600 & TVS600A Command Reference
2–123
None
INPut:FILTer
INPut:FILTer?
Sets or queries whether the low-pass analog noise filter is on or off for the
specified input channel. INPut:FILTer must be off to use the full bandwidth of
the waveform analyzer. To set the limit frequency of the low pass filter, use the
command INPut:FILTer:FREQuency.
INPut<n>:FILTer[:LPASs][:STATe] <boolean>
INPut<n>:FILTer[:LPASs][:STATe]?
<n> (Channel Number)
1
Query response
1
2
3
4
Not applicable
1
The input channel number <n> is used by the commands INPut<n> and VOLTage<n>.
If you omit <n>, the default is channel 1.
<boolean> Query response
<NRf>
1 or ON
0 or OFF
<NF1>
1
0
0 (OFF)
Execution Error –241, “Hardware missing”
Attempted to program INPut3 or INPut4 on an
instrument configured with two channels.
None
Command: INP1:FILT ON
Query: INP1:FILT?
Response: 1
Related Commands
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples