CALCulate Subsystem
TVS600 & TVS600A Command Reference
2–71
Execution Error –141, “Invalid character data,
CALC1:WMP FTM”
Attempted to set method to an illegal value.
None
Command: CALC1:WMP:HMET ABS
Query: CALC1:WMP:HMET?
Response: ABS
CALCulate:WMParameter:LMEThod
CALCulate:WMParameter:HIGH
CALCulate:WMParameter:LOW
CALCulate:WMParameter:LOW?
Sets or queries the low (most negative) level used for time and amplitude
measurements. This setting is used only when CALC:WMP:LMEThod is set to
ABSolute. For more information, refer to the discussion of CALC:WMP:HME-
Thod. The units are those currently used for vertical magnitude.
CALCulate<n>:WMParameter:LOW <low>
CALCulate<n>:WMParameter:LOW?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
<low> (Absolute LOW level) Query response
<NRf> <NR3>
0.0E+0