Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
SYSTem Subsystem
TVS600 & TVS600A Command Reference
2–205
SYSTem:PROTect is enabled, set to ON, when you reset (*RST) or power up the
instrument.
SYSTem:PROTect
SYSTem:PROTect?
<boolean> Query response
<NRf>
1 or ON
0 or OFF
<NR1>
1
0
1
None
None
Command: SYST:PROT ON
Query: SYST:PROT?
Response: 1
INPut:PROTection:STATe
SYSTem:SECurity:IMMediate
TEST:LOG:CLEar
*PUD
SYSTem:SECurity:IMMediate
Immediately destroys all measurement data and stored instrument settings. The
instrument writes over measurement data in acquisition memory three times with
a fixed bit pattern. Current settings are initialized to their *RST or factory default
values but critical calibration constants are retained. Set SYSTem:PROTect to
OFF before using the command SYSTem:SECurity:IMMediate.
SYSTem:SECurity:IMMediate
None
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands
Syntax
Parameters