Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
SYSTem Subsystem
2–194
TVS600 & TVS600A Command Reference
Queue. The OFF mode is appropriate when you have a computer connected to
the RS-232 connector. When ON, messages are immediately displayed. The ON
mode is most appropriate when a simple display terminal is connected to the
front panel RS-232. Refer to the TVS600 & TVS600A Series Waveform Analyzers
User Manual for the factory setting.
When you set :SERial:ERESponse, the new value is stored in nonvolatile RAM
and is recalled at power up. However, the error response mode is not saved with
stored settings (*LRN) and it is not reset to a default value by *RST.
SYSTem:COMMunicate:SERial:ERESponse <boolean>
SYSTem:COMMunicate:SERial:ERESponse?
<boolean> Query response
<NRf>
1 or ON
0 or OFF
<NR1>
1
0
Not applicable
None
None
Command: SYST:COMM:SER:ERES OFF
Query: SYST:COMM:SER:ERES?
Response: 0
SYSTem:COMMunicate:SERial:PRESet:RAW
SYSTem:COMMunicate:SERial:PRESet:TERMinal
SYSTem:COMMunicate:SERial:LBUFfer
SYSTem:COMMunicate:SERial:LBUFfer?
Sets or queries the state of the character buffer for the front panel RS-232 port.
When ON, all input characters are queued until a newline (^J) or return (^M)
character is received; then the buffer contents are released to the command
parser. The following line editing commands let you modify the input before
entering a newline or return character:
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands