VOLTage Subsystem
2–306
TVS600 & TVS600A Command Reference
<lower> (Lower limit of voltage range)
1
Query response
<NRf> <NR3>
1
The default multiplier for the <lower> parameter is V for volts. You can also use MV
for millivolts.
–0.5E+0
Execution Error –222, “Data out of range”
Attempted to set low pass frequency to an illegal value.
Execution Error –241, “Hardware missing”
Attempted to program VOLTage3 or VOLTage4 on an instrument configured
with only two channels.
VOLTage:RANGe[:UPPer] and VOLTage:RANGe:LOWer operate as a pair.
Changing one will not cause the other to change. However, a change in the
voltage range will change VOLTage:RANGe:PTPeak and VOLT-
age:RANGe:OFFSet so they describe the new voltage range.
VOLTage:RANGe:LOWer is not returned in the response to *LRN? because the
voltage range is returned as VOLT:RANG:PTP and VOLT:RANG:OFFS.
Command: VOLT1:RANG:LOW -2
Query: VOLT1:RANG:LOW?
Response: -2.0E+0
VOLTage:RANGe[:UPPer]
VOLTage:RANGe:OFFSet
VOLTage:RANGe:PTPeak
Reset Value
Errors and Events
Dependencies
Examples
Related Commands