CALCulate Subsystem
TVS600 & TVS600A Command Reference
2–83
CALCulate:WMParameter:SLOPe
CALCulate:WMParameter:SLOPe?
Sets or queries the slope of the edge desired for delay waveform measurements.
This setting only affects DELay measurements. All other edge measurements
determine the edge slope implicitly from the measurement specified. For
example, RTIMe uses positive slope.
CALCulate[1]:WMParameter:SLOPe <slope>
CALCulate[1]:WMParameter:SLOPe?
CALCulate[2]:WMParameter:SLOPe <slope>
CALCulate[2]:WMParameter:SLOPe?.
CALCulate[3]:WMParameter:SLOPe <slope>
CALCulate[3]:WMParameter:SLOPe?.
CALCulate[4]:WMParameter:SLOPe <slope>
CALCulate[4]:WMParameter:SLOPe?
<slope> Query response
NEGative
POSitive
NEG
POS
None (default vertical units)
POS
Execution Error –141, “Invalid character data”
Attempted to set SLOPe to an illegal value.
None
Command:
CALC1:WMP:SLOP NEG
Query: CALC1:WMP:SLOP?
Response: NEG
CALCulate:WMParameter:EDGE
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands