AADVance Subsystem
2–4
TVS600 & TVS600A Command Reference
AADVance
AADVance?
Sets or queries the state of the auto-advance acquisition mode. In the auto-ad-
vance mode the waveform analyzer acquires a sequence of data records for each
active channel. The delay between one acquisition record and the next one is
very short and is due only to the minimal re-arm time and any trigger holdoff
you set. Use the command AADVance:COUNt to set the number of records to
acquire. You cannot use acquisition modes average, envelope, or peak detect (set
using [SENSe:]AVERage command) when using the auto-advance acquisition
mode.
The auto-advance mode of acquisition affects all enabled channels (XTIM:VOLT
<n>). You cannot acquire one channel in the auto-advance mode and acquire
another with the normal acquisition mode.
[SENSe:]AADVance[:STATe] <boolean>
[SENSe:]AADVance[:STATe]?
<boolean> Query response
<NRf>
1 or ON
0 or OFF
<NF1>
1
0
0
None
Enabling auto-advance acquisition disables [SENSe:]AVERage acquisition,
ENVelope or SCALar type.
Command: AADV ON
Query: AADV?
Response: 1
AADVance:COUNt
AVERage
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands