Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
FORMat Subsystem
TVS600 & TVS600A Command Reference
2–99
FORMat:TRACe:REF
FORMat:TRACe:REF?
TVS600A Models Only
This command sets/queries the type of encoding used to transfer a Reference
trace through word serial and RS232 external interfaces.
The number of significant digits returned by an ASCII data query is not
selectable. A setting of ASCii, 0 selects a format with a varying number of
significant digits depending on the value of the data point. Refer to the TVS600
& TVS600A Series Waveform Analyzers User Manual for a definition of the data
formats for <type>.
This command has no effect on FDC data transfers which are always REAL,32.
FORMat[:DATA]:TRACe:REF <type>[,<length>]
FORMat[:DATA]:TRACe:REF?
<type>,<length> Query response
ASCii[,0]
REAL,32
ASC,0
REAL,32
None
ASC,0
Command Error –141, “Invalid character data” or
Execution Error –224, “Illegal parameter value”
Attempted to set format to an illegal value.
None
Command: FORM:TRAC:REF REAL,32
Query: FORM:TRAC:REF?
Response: REAL,32
TRACe[:DATA]?
TRACe[:DATA]:AATS?
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands