Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALibration Subsystem
2–90
TVS600 & TVS600A Command Reference
CALibration:PROBe[n]
CALibration:RESults? (Query Only)
Returns the results code for the last calibration performed. The result code is the
numeric identifier for the first function to fail. A value of zero is returned when
there are no failures and a value of –1 indicates a calibration is in progress.
Calibration functions have numeric identifiers in the range 2000 to 2999.
Because this query is not synchronized with instrument operations, you may
want to use *WAI, *OPC, or *OPC? to wait for the CAL_OPC pending flag to
clear before sending the CAL:RES? query.
CALibration:RESults[:CODE]?
Query response
Not applicable <NR1>
–1 Calibration in progress
0 No failures
2000 N 2999
Not applicable
None
None
Query: CAL:RES?
Response: 0
CALibration
CALibration:RESults:VERBose?
Related Commands
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands