IEEE 488.2 Common Commands
TVS600 & TVS600A Command Reference
2–327
*TST? (Query Only)
Initiates an internal self-test and returns a failure code. If more than one failure
occurred, only the test number of the first failure is reported. A value of zero
indicates there were no failures. This query is only valid when looping is
disabled. Detailed test results are available with the TEST:RESult:VERBose?
command.
*TST? and TEST? perform the same function.
*TST?
<failure_code> Query response
Not applicable <NR1>
0 (no failure)
1000 – 1999 (Self-test failure)
2000 – 2999 (Calibration failure)
Not applicable
Execution Error –221, “Settings conflict”
Executed *TST? when TEST:CONTrol:LOOP is ON.
None
Query: *TST?
Response: 0
TEST
TEST:RESults:VERBose?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples