Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
IEEE 488.2 Common Commands
TVS600 & TVS600A Command Reference
2–321
*RCL
Recalls the specified instrument setting from one of the ten non-volatile memory
locations in the waveform analyzer. You save settings to these locations with the
*SAV command.
*RCL <location>
<location> Query response
<NRf>
1 to 10
Not applicable
Not applicable
Execution Error –221, “Settings conflict”
Attempted to recall an instrument setting from an empty memory location.
Execution Error –224, “Illegal parameter value”
Attempted to recall an instrument setting from an illegal memory location.
None
Command: *RCL 4
SYSTem:SET
*LRN?
*SAV
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands