CALibration Subsystem
TVS600 & TVS600A Command Reference
2–89
CALibration:PROBe:RESults?
TVS600A Models Only
Returns the calibration status of connected level 2 probes.
1. A value of one (1) returned by the query indicates an initialized (uncali-
brated) probe.
2. A value of zero (0) indicates a calibrated probe.
3. A value of minus one (–1) indicates a probe calibration failure.
4. A value of minus two (–2) indicates a probe calibration failed because the
probe was not connected to the calibration source.
5. A value of minus three (–3) indicates a calibration failed because an
unsupported probe or no probe was installed.
CALibration:PROBe[1]:RESults?
CALibration:PROBe2:RESults?
CALibration:PROBe3:RESults?
CALibration:PROBe4:RESults?
<calibration–status> Query response
Not applicable <NR1>
0
1
–1
–2
–3
Not applicable
Execution Error –241, “Hardware missing” Attempted to query PROBe3 or
PROBe4 when the instrument is configured with two channels.
None
Command: CAL:PROB1
Query: CAL:PROB1?
Response: 0
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples