CALCulate Subsystem
TVS600 & TVS600A Command Reference
2–23
CALCulate:AAMList:STATe
CALCulate:AAMList:STATe?
Sets or queries whether to perform waveform measurement(s) on acquisition
records captured with auto-advance acquisition. You specify Auto Advance
measurements with the CALCulate:AAMList command.
CALCulate<n>:AAMList:STATe <boolean>
CALCulate<n>:AAMList:STATe?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
<boolean> Query response
<NRf>
1 or ON
0 or OFF
<NF1>
1
0
0
None
The setting for :STATe is ignored when you use the command CALC:PATH:EX-
PRession.
Command: CALC:AAML:STAT ON
Query: CALC:AAML:STAT?
Response: 1
CALCulate:AAMList