Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALCulate Subsystem
TVS600 & TVS600A Command Reference
2–43
Response: 10.01E+6
CALCulate:FILTer:FREQuency:STARt
CALCulate:FILTer:FREQuency:CENTer
CALCulate:FILTer:FREQuency:SPAN
CALCulate:FILTer:FREQuency:TWIDth
CALCulate:FILTer:FREQuency:TWIDth?
Sets or queries the slope of roll off for the post-acquisition filter. The filter slope
is specified as a ratio of the absolute transition width (in Hz) to Nyquist
frequency defined as 1 / (2 × <sample interval>). A low value for transition
width produces a steep slope for the filter while a large value produces a gradual
slope or roll off. You can also specify :TWIDth in percent.
If you enter too small a value for :TWIDth, the function will generate too many
filter coefficients resulting in an execution error and termination of the filter
process. For more information, refer to the filter discussion in the TVS600 &
TVS600A Series Waveform Analyzers User Manual.
CALCulate<n>:FILTer[:GATE]:FREQuency:TWIDth <width>
CALCulate<n>:FILTer[:GATE]:FREQuency:TWIDth?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
<width> (Rate of transition in percent) Query response
<NRf>
0.0004 N 1.0
MINimum
MAXimum
<NR2>
0.0004
1.0
0.1
Related Commands
Syntax
Parameters
Reset Value