IEEE 488.2 Common Commands
2–314
TVS600 & TVS600A Command Reference
*ESE
*ESE?
Sets or queries the Event Status Enable Register (ESER). The ESER determines
which events in the Standard Event Status Register can set the Event Status bit
(bit 5) in the Status Byte Register. The bits in the ESER correspond to the bits in
the Standard Event Status Register, which are defined in Table 2–6 on page
2–315.
The STATus:PRESet command does not change the Event Status Enable
Register. For more information on the Status and Events reporting system, refer
to the TVS600 & TVS600A Series Waveform Analyzers User Manual.
*ESE <mask>
*ESE?
<mask> Query response
<NRf> | <Non-decimal number>
0 ≤ N ≤ #HFF
<NR1>
Not applicable
None
None
Command: *ESE #H40
This command enables only the User Request bit (bit6) in the
Standard Event Status Register. No other events can set the Event
Status bit in the Status Byte Register.
Query: *ESE?
Response: 64
*ESR?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands