Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
OUTPut Subsystem
2–144
TVS600 & TVS600A Command Reference
OUTPut:REFerence:FUNCtion
OUTPut:REFerence:FUNCtion?
Sets or queries which reference signal is output on the connector REFERENCE
OUTPUT. The reference signal is intended for use during instrument verifica-
tion. You can direct either of two signals to the REFERENCE OUTPUT: the
internal 10 MHz reference clock with a 1 V peak-to-peak amplitude or the 8.0 V
DC level. Use OUTPut:REFerence to enable the signal output. The reference
signal output is disabled when you start acquisition or reset the instrument.
OUTPut:REFerence:FUNCtion <function>
OUTPut:REFerence:FUNCtion?
<function> Query response
CLOCk
VOLTage
CLOC
VOLT
CLOC
Execution Error –141, “Invalid character data”.
Attempted to set the calibration reference function to an illegal value.
None
Command: OUTP:REF:FUNC VOLT
Query: OUTP:REF:FUNC?
Response: VOLT
OUTPut:REFerence
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands