TRIGger:PULSe Subsystem
2–270
TVS600 & TVS600A Command Reference
TRIGger:PULSe:CLASs
TRIGger:PULSe:CLASs?
Sets or queries the class of pulse triggering to use for the next acquisition. The
trigger classes are as follows:
H GLITch — the trigger event occurs when the instrument detects a pulse that
is either narrower or wider than a specified duration or width.
H TIMEout —the trigger event occurs when the pulse width timer expires on
pulses wider than the specified duration
H WIDTh — the trigger event occurs when the instrument detects a pulse that
is either inside or outside a specified duration or width.
This command has effect only when you set TRIGger:TYPE to PULSe. The
defined alias for the SCPI trigger :SEQuence[1] is A.
TRIGger[:A]:PULSe:CLASs <pulse_class>
TRIGger[:A]:PULSe:CLASs?
<pulse_class> Query response
GLITch
TIMEout
1
WIDTh
GLIT
TIME
WIDT
1
TVS600A models only
GLITch
Execution Error –224, “Illegal parameter value”
Attempted to set class to an illegal value.
None
Command: TRIG:PULS:CLAS WIDTH
Query: TRIG:PULS:CLAS?
Response: WIDT
TRIGger:TYPE
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands