Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
TRIGger:PULSe Subsystem
2–278
TVS600 & TVS600A Command Reference
TRIGger:PULSe:WIDTh:HLIMit
TRIGger:PULSe:WIDTh:HLIMit?
Sets or queries the high or longest valid pulse width for to qualify for pulse
triggering. After a pulse on the trigger source begins, it cannot take longer than
:WIDTh:HLIMit to cross the TRIGger:PULSe:THReshold level. HLIMit must
be longer than :LLIMit. To activate pulse width triggering, set TRIGger:TYPE to
PULSe and TRIGger:PULSe:CLASs to WIDTh.
TRIGger[:A]:PULSe:WIDTh:HLIMit <hlimit>
TRIGger[:A]:PULSe:WIDTh:HLIMit?
<hlimit>
1
Query response
<NRf>
MIN
MAX
<NR3>
2.0E–9
1.0E+0
1
The default for the parameter <hlimit> is seconds (S). You can also use the
multipliers MS for milliseconds, US for microseconds, NS for nanoseconds, and PS
for picoseconds.
4.0E–9
Execution Error –222, “Data out of range”
Attempted to set high limit to an illegal value.
None
Command: TRIG:PULS:WIDT:HLIM 1.0
Query: TRIG:PULS:WIDT:HLIM?
Response: 1.0E+0
TRIGger:TYPE
TRIGger:PULSe:CLASs
TRIGger:PULSe:WIDTh:LLIMit
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands