FUNCtion and DATA Subsystems
2–108
TVS600 & TVS600A Command Reference
FUNCtion:CONCurrent
FUNCtion:CONCurrent?
Sets or queries whether more than one sense function can be enabled at a time.
When you set FUNCtion:CONCurrent from OFF to ON, the state of individual
functions does not change. All disabled functions remain off and if one function
was enabled, it remains on. While FUNCtion:CONCurrent is ON, you can
enable any and all provided functions.
When FUNCtion:CONCurrent is OFF, and you define a new function with
FUNCtion[:ON], the specified function is enabled and all others are disabled.
[SENSe:]FUNCtion:CONCurrent <boolean>
[SENSe:]FUNCtion:CONCurrent?
<boolean> Query response
<NRf>
1 or ON
0 or OFF
<NR1>
1
0
1
None
Setting FUNCtion:CONCurrent from ON to OFF will enable XTIM:VOLT 1 and
disable all other functions.
Command: FUNC:CONC ON
Query: FUNC:CONC?
Response: 1
FUNCtion[:ON]
FUNCtion:OFF
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands