SWEep Subsystem
TVS600 & TVS600A Command Reference
2–181
Execution Error –241, “Hardware missing”
Attempted to set time interval to 200E–12 when the instrument is a TVS621A,
TVS641A; or to set time interval to 400E-12 when the instrument is a TVS621A
or TVS641A.
Changing SWEep:TINTerval changes SWEep:TIME. The relation is
SWEep:TINTerval = SWEep:TIME / SWEep:POINts.
Changing SWEep:TIMTerval to < 100 ns changes SWEep:POINts to 15,000 if
previously set to 30,000. The acquisition mode changes from extended real time
to real time mode below 100 ns/sample.
Command: SWE:TINT 1E-6
Query: SWE:TINT?
Response: 1.0E-6
SWEep:POINts
SWEep:TIME
Dependencies
Examples
Related Commands