Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
TRIGger:PULSe Subsystem
TVS600 & TVS600A Command Reference
2–275
TRIGger:PULSe:THReshold
TRIGger:PULSe:THReshold?
Sets or queries the voltage threshold used for pulse triggering. The threshold is
the voltage level that the pulse trigger source must cross to mark the beginning
and end of a pulse. To set positive or negative pulse polarity use the command
TRIG:PULS:WIDT:POL or TRIG:GLIT:WIDT:POL, depending on the selected
type of pulse triggering.
TRIGger[:A]:PULSe:THReshold <level>
TRIGger[:A]:PULSe:THReshold?
<level>
1
Query response
<NRf>
Internal DC Coupled:
absolute maximum
2
: –200.0 N 200.0
limited by ± VOLT:RANG:PTP
steps 0.002 * VOLT:RANG:PTP
MINimum
2
MAXimum
2
<NR3>
VOLT:RANG:OFFS ± VOLT:RANG:PTP
VOLT:RANG:OFFS – VOLT:RANG:PTP
VOLT:RANG:OFFS + VOLT:RANG:PTP
1
The default for the parameter <level> is V for volts. You can also use the multipliers
MV for millivolts and UV for microvolts.
2
When you connect a probe, the maximum limits increase just as the vertical range
increases for the input. For example, with a 1 V vertical range, connecting a 10X
probe increases the vertical range and trigger level range to 10 V.
0.0E+0
Execution Error –222, “Data out of range”
Attempted to set threshold to an illegal value.
None
Command: TRIG:PULS:THR 1.0
Query: TRIG:PULS:THR?
Response: 1.0E+0
TRIGger:TYPE
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands