Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALCulate Subsystem
2–72
TVS600 & TVS600A Command Reference
Execution Error –222, “Data out of range”
Attempted to set LOW to an illegal value.
None
Command: CALC1:WMP:LOW -10.0
Query: CALC1:WMP:LOW?
Response: -10.0E+0
CALCulate:WMParameter:HIGH
CALCulate:WMParameter:LMEThod
CALCulate:WMParameter:LMEThod
CALCulate:WMParameter:LMEThod?
Sets or queries the method for calculating the LOW (most negative) value for
time and amplitude waveform measurements.
ABSolute — specifies that LOW is set to the value of CALCulate:WMPara-
meter:LOW.
AUTO — selects the MODE method of setting LOW when the histogram
function is able to detect a consistent level below MID. Otherwise, the
PEAK method is used. The AUTO method is effective when you are not
certain what type of waveform to expect.
MODE — selects the level for LOW based on a peak histogram function
which looks for a greater than average number of data points at a level below
MID. This method is useful to ignore spurious peaks on a digital logic
waveform, such as a TTL clock signal.
PEAK — specifies that LOW is set to the lowest amplitude data point in the
acquisition record. This method is useful for a sinewave or triangle
waveform.
CALCulate<n>:WMParameter:LMEThod <method>
CALCulate<n>:WMParameter:LMEThod?
Errors and Events
Dependencies
Examples
Related Commands
Syntax