Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
TRIGger:TRANsition Subsystem
TVS600 & TVS600A Command Reference
2–295
TRIGger:TRANsition:RUNT:QUALify
TRIGger:TRANsition:RUNT:QUALify?
TVS600A Models Only
Sets or queries the type of time qualification for RUNT transition triggering.
OFF means there is no time qualification. GT means the trigger occurs when the
transition time is greater than the specified time.
:A is a predefined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is TRANsition, and
TRIGger:TRANsition:CLASs is RUNT.
TRIGger[:A]:TRANsition:RUNT:QUALify <type>
TRIGger[:A]:TRANsition:RUNT:QUALify?
<type> Query response
GT
OFF
GT
OFF
None
OFF
Command Error –141, “Invalid character data”
Attempted to set QUALify to an illegal value.
None
Command: TRIG:TRAN:RUNT:QUAL GT
Query: TRIG:TRAN:RUNT:QUAL?
Response: GT
TRIGger:TYPE
TRIGger:TRANsition:CLASs
TRIGger:TRANsition:TIME
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands