Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
TRIGger:LOGic Subsystem
2–264
TVS600 & TVS600A Command Reference
TRIGger:LOGic:PATTern:QUALify
TRIGger:LOGic:PATTern:QUALify?
TVS600A Models Only
Sets or queries the type of time qualification performed when pattern logic
triggering is selected. OFF means that there is no time qualification. LT means
that the trigger occurs when the pattern is true less than the specified width. GT
means that the trigger occurs when the pattern is true greater than the specified
width.
:A is a predefined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is set to LOGic and
TRIGger:LOGic:CLASs is PATTern.
TRIGger[:A]:LOGic:PATTern:QUALify <type>
TRIGger[:A]:LOGic:PATTern:QUALify?
<type> Query response
GT
LT
OFF
GT
LT
OFF
None
OFF
Command Error –141, “Invalid character data”
Attempted to set QUALify to an illegal value.
None
Command: TRIG:LOG:PATT:QUAL GT
Query: TRIG:LOG:PATT:QUAL?
Response: GT
TRIGger:TYPE
TRIGger:LOGic:CLASs
TRIGger:LOGic:PATTern:WIDTh
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands