Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALCulate Subsystem
2–66
TVS600 & TVS600A Command Reference
CALCulate:WMParameter:GATE:STARt:RELative
CALCulate:WMParameter:GATE:STARt:RELative?
TVS600A Models Only
Sets or queries the relative starting location for gated waveform measurements.
This setting is ignored unless CALC:WMP:GATE is set to ON and
CALC:WMP:GATE:METH is set to REL.
CALCulate[1]:WMParameter:GATE:STARt:RELative <ratio>
CALCulate[1]:WMParameter:GATE:STARt:RELative?
.
.
CALCulate4:WMParameter:GATE:STARt:RELative <ratio>
CALCulate4:WMParameter:GATE:STARt:RELative?
<ratio> Query response
<NRf> <NR3>
PCT
0.0
None
None
Command:
CALC1:WMP:GATE:STAR:REL 0.0
Query: CALC1:WMP:GATE:STAR:REL?
Response: 0.0E+0
CALCulate:WMParameter:GATE
CALCulate:WMParameter:GATE:METHod
CALCulate:WMParameter:GATE:STOP:RELative
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands