Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALCulate Subsystem
2–62
TVS600 & TVS600A Command Reference
CALCulate:WMParameter:EDGE
CALCulate:WMParameter:EDGE?
Sets or queries the index of the edge desired for cross and delay waveform
measurements. This setting affects FTIMe, RTIMe, CROSs, NCRoss, PCRoss
DELay, and COPulse measurements. A positive <edge> value searches from the
beginning of the waveform record. A negative <edge> value searches from the
end of the waveform record. Zero <edge> is the last edge in the record.
Measurement functions with implcitly defined slope (FTIMe, RTIMe, NCRoss,
PCRoss) or explicitly defined slope (DELay) only count edges of the proper
slope. CROSs and COPulse count edges of either slope.
CALCulate[1]:WMParameter:EDGE <edge>
CALCulate[1]:WMParameter:EDGE?
.
.
CALCulate4:WMParameter:EDGE <edge>
CALCulate4:WMParameter:EDGE?
<edge> Query response
<NRf> <NR1>
None (default vertical units)
1
Execution Error –222, “Data out of range”
Attempted to set EDGE to an illegal value.
None
Command: CALC1:WMP:EDGE 4
Query: CALC1:WMP:EDGE?
Response: 4
CALCulate:WMParameter:LOW
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands