OUTPut Subsystem
2–138
TVS600 & TVS600A Command Reference
OUTPut:ECLTrg<n>
OUTPut:ECLTrg<n>?
Sets or queries whether the instrument should drive (source) the specified
VXIbus ECL trigger line when a trigger event occurs. You can choose inverted
output polarity on the ECL trigger lines.
The waveform analyzer lets you to define the source to drive each VXIbus ECL
trigger line. You can source and sense an ECL trigger line.
OUTPut:ECLTrg<n>[:STATe] <boolean>
OUTPut:ECLTrg<n>[:STATe]?
<n> (number of VXIbus trigger signal) Query response
<NR1>
0 or 1
<NR1>
<boolean> Query response
<NRf>
1 or ON
0 or OFF
<NR1>
1
0
0
None
None
Command: OUTP:ECLT0 ON
Query: OUTP:ECLT0?
Response: 1
OUTPUt:ECLTrg<n>:SOURce
OUTPUt:ECLTrg<n>:POLarity
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands