TRIGger:LOGic Subsystem
TVS600 & TVS600A Command Reference
2–267
TRIGger:LOGic:THReshold
TRIGger:LOGic:THReshold?
TVS600A Models Only
Sets or queries the trigger threshold for each channel when logic triggering is
selected.
:A is a predefined alias for :SEQuence[1].
The minimum and maximum values for threshold are VOLT:RANG:OFFS +/–
VOLT:RANG:PTP (in VOLT:RANG:PTP * 0.002 steps).
This command only has effect when TRIGger:TYPE is LOGic.
TRIGger:LOGic:THReshold1 is associated with INPut1, etc.
TRIGger[:A]:LOGic:THReshold[1] <level>
TRIGger[:A]:LOGic:THReshold[1]?
TRIGger[:A]:LOGic:THReshold2 <level>
TRIGger[:A]:LOGic:THReshold2?
TRIGger[:A]:LOGic:THReshold3 <level>
TRIGger[:A]:LOGic:THReshold3?
TRIGger[:A]:LOGic:THReshold4 <level>
TRIGger[:A]:LOGic:THReshold4?
<level> Query response
<NRf>
MINimum
MAXimum
<NR3>
<See comments>
<See comments>
V (default), MV, or UV
0.0V
Execution Error –222, “Data out of range”
Attempted to set threshold to an illegal value.
Execution Error –241, “Hardware missing”
Attempted to program THReshold3 or THReshold4
when the instrument is configured with two channels.
None
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies