Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALCulate Subsystem
TVS600 & TVS600A Command Reference
2–25
CALCulate:DATA:PREamble? (Query Only)
Returns the data preamble for acquisition record processing and measurement
functions. Results are returned only when pending acquisitions and calculations
are complete. The default format of the returned data is ASCII. However, you
can set the data format with the FORMat:CALCulate command. The synchroniz-
ing commands *WAI, *OPC, and *OPC? are not required unless you wish to
synchronize the transfer differently.
The CALCulate and TRACe subsystems use consistent naming, such that
CALC1:DATA? is equivalent to TRAC? CALC1.
CALCulate<n>:DATA:PREamble?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
Not applicable
Execution Error –230, “Data corrupt or stale”
Attempted to query the preamble for data that is invalid, incomplete or stale.
None
Query: CALC1:DATA:PRE?
Response: (DIF Expression)
[SENSe:]DATA?
[SENSe:]DATA:PREamble?
CALCulate:DATA?
TRACe?
TRACe:PREamble?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands