CALCulate Subsystem
2–46
TVS600 & TVS600A Command Reference
CALCulate:IMMediate
CALCulate:IMMediate?
Sets or queries whether the specified CALCulate block will reprocess SENSe
data without reacquiring new data. The calculation is performed immediately
and, if the query form is sent, the resulting data is returned afterward. The query
form CALC:IMM? is semantically equivalent to CALC:IMM;DATA?. The
format of the resulting data is determined by the FORMat subsystem.
This command sets the ACQ_OPC pending flag. *WAI, *OPC, or *OPC? may
be used to synchronize the command.
CALCulate<n>:IMMediate
CALCulate<n>:IMMediate?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
<data> Query response
Not applicable Defined by FORMat:CALCulate
Not applicable
The only errors are those associated with the defined measurements or calcula-
tions. Most calculation errors appear only at execution time rather than when you
define a measurement list. The most common errors are Execution Error –260,
“Expression error” and the Questionable Event errors: 2090, “Calculate1
questionable”; 2100, “Calculate2 questionable”; 2110, “Calculate3 question-
able”; 2120, “Calculate4 questionable”.
None
Command: CALC1:IMM
Query: CALC1:IMM?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples