Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
TEST Subsystem
2–210
TVS600 & TVS600A Command Reference
TEST
TEST?
Executes all internal self-tests once. If a failure occurs, the test will stop. The
query form returns the number of the first test that fails or zero if there were no
failures. The command form executes the same tests but does not return a result
value.
Use the command *WAI, *OPC, or *OPC? to delay execution of the TEST?
command until the self test completes. Completion of the tests clears the
TESTing flag in the Operation Status register.
Upon a failure, you should refer the instrument to your Service provider for
repair. For a listing of the failure codes, refer to the TVS600 Series Waveform
Analyzers Service Manual.
TEST[:ALL]
TEST[:ALL]?
None Query response
Not applicable <NR1>
0 (no failure)
1000 – 1999 (self test failure)
Not applicable
None
Command: TEST
Query: TEST?
Response: 0
TEST:RESults?
TEST:RESults:VERBose?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands