CALibration Subsystem
2–86
TVS600 & TVS600A Command Reference
CALibration
CALibration?
Executes all self-calibration functions. If a failure occurs, calibration will
immediately stop. The query form runs the self-calibration functions and returns
the numeric identifier of the first failed function. A value of zero indicates there
were no failures. The command form executes the functions but returns no
results code.
Calibration functions have numeric identifiers in the range 2000 to 2999.
CALibration sets the CAL_OPC pending flag. Use the command *WAI, *OPC,
or *OPC? to synchronize the response with completion of self calibration.
To compensate probes, use the command OUTPut:PCOMpensate which enables
the probe calibration signal to the front panel.
CALibration[:ALL]
CALibration[:ALL]?
<boolean> Query response
Not applicable <NR1>
0 No failures
2000 N 2999
Not applicable
None
None
Command: CAL
Query: CAL?
Response: 0
CALibration:RESults?
CALibration:RESults:VERBose?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands