CALCulate Subsystem
2–82
TVS600 & TVS600A Command Reference
CALCulate:WMParameter:RMEThod
CALCulate:WMParameter:RMEThod?
Sets or queries the method for calculating the reference (high, middle, low)
values for time and amplitude measurements. Selecting ABSolute lets you set
the measurement parameters at absolute vertical levels. RELative lets you set
them as a ratio or percentage of the parameter to the vertical amplitude of the
record data.
CALCulate<n>:WMParameter:RMEThod <method>
CALCulate<n>:WMParameter:RMEThod?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
<method> (Sets reference method) Query response
ABSolute
RELative
ABS
REL
REL
Execution Error –141, “Invalid character data”
Attempted to set method to an illegal value.
None
Command: CALC1:WMP:RMET ABS
Query: CALC1:WMP:RMET?
Response: ABS
CALCulate:WMParameter:HREFerence
CALCulate:WMParameter:LREFerence
CALCulate:WMParameter:MREFerence
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands