Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
TRIGger:TRANsition Subsystem
2–302
TVS600 & TVS600A Command Reference
TRIGger:TRANsition:TIME
TRIGger:TRANsition:TIME?
TVS600A Models Only
Sets or queries the transition qualification time.
:A is a predefined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is TRANsition and
TRIGger:TRANsition:QUALify is LT or GT.
TRIGger[:A]:TRANsition:TIME <time>
TRIGger[:A]:TRANsition:TIME?
<time> Query response
<NRf>
MINimum
MAXimum
<NR3>
2.0E–9
1.0E+0
S (default), MS, US, or PS
2.0E–9
Execution Error –222, “Data out of range”
Attempted to set time to an illegal value.
None
Command: TRIG:TRAN:TIME 1.0
Query: TRIG:TRAN:TIME?
Response: 1.0E+0
TRIGger:TYPE
TRIGger:TRANsition:QUALify
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands